TECHNICAL SPECIFICATIONS
The unit is designed in such a way that it requires little or no oil.
The pump is equipped with a provision preventing fluid flowing to the pump if the collection jar is full (overflow control valve).
The unit should provide intermittent and continuous suction. Continuous is limited to 30 minutes after which the unit should be given 30 minutes to cool down.
The unit should be provided with two (2) reusable collection jars.
The unit should have maintenance free operating.
A gauge indicating the negative pressure generated by the unit (preferably in kPa or mmHg).
A vacuum control and on/off switch are present on the casing.
The flow rate at maximum vacuum is ≥ 40 L/m (2.4 m³/h).
A suction tube of at least 1.5 meters is included.
A hydrophobic bacterial filter is included in the airline to the pump.
The unit has user-adjustable suction between 0 and at least 600 mmHg (80 kPa).
The suction tube is autoclavable at 121°C.
The unit is equipped with a foot switch for control of the suction.
During operating the unit’s noise level does not exceed 60.1 dB.
The unit should be mounted on a wheeled, stable, portable stand.
The device should have a sturdy, shock resistant casing.
The body is to be constructed of plastic in accordance with the latest European standards.
Power requirements: 100 – 240 Volts – 50/60 Hz (not necessarily in a single unit).
The collection jar is made of shatterproof polycarbonate.
The stand is equipped with 360°, anti-static, swivel castors.
A least two (2) castors should be lockable.
The design of the device should be such that it is easy to clean/disinfect by avoiding difficult-to-reach areas and having smooth surfaces.